Content » Vol 51, Issue 4

Original report

Quantifying upper extremity performance with and without assistance of a soft-robotic glove in elderly patients: A kinematic analysis

Anne van Ommeren, Bob Radder, Anke Kottink, Jaap Buurke, Gerdienke Prange-Lasonder, Johan Rietman
Roessingh Research and Development, 7522 AH Enschede, The Netherlands. E-mail: a.vanommeren@rrd.nl
DOI: 10.2340/16501977-2535

Abstract

Objective: To explore the direct influence of a soft-robotic glove on movement duration and movement execution in elderly people with decreased hand function during a reach-and-grasp task.
Design: Cross-sectional study.
Patients: Eight subjects, aged 55+ years, with decreased hand function.
Methods: The direct effect of the glove was explored using kinematic analysis during a reach-and-grasp task with a light (100 g) and heavy (1,000/2,500 g) cylindrical object, performed with and without the soft-robotic glove.
Results: There was no difference in total movement time between performance with and without the glove. With the glove, the relative time needed to transport the heavy object was shorter, while the relative time needed to grasp the heavy object was longer. In addition, transporting light objects involved a lower peak velocity and larger elbow extension, and grasping the object involved a larger hand opening compared with without glove.
Conclusion: As expected, no positive influence of the soft-robotic glove was found on total movement duration in elderly subjects. The influence of the glove on movement execution varied with movement phase. The positive and negative effects found may be due to a perceived confidence while carrying heavy objects with the glove, or compensation for loss of sensation, respectively. This information can be used to improve the glove design.

Lay Abstract

The function of the ageing hand decreases as result of loss of muscle mass or age-related diseases, such as osteoarthritis or rheumatoid arthritis. This loss of func­tion results in limitations in performing activities of daily living, such as carrying heavy objects, drinking and eating. A wearable soft-robot glove (the ironHand) has been developed to support grip strength during daily life. This study assessed the effect of the ironHand on movement execution in 8 elderly people with decreased hand function due to age-related diseases. The influence of the glove on movement execution in elderly subjects varied with movement phase. Grasping of a heavy object took relatively longer, while its transport phase was relatively shorter, with the glove, compared with without the glove. These results provide insight into how a soft-robotic glove influences movement, both in a positive and a negative sense. This information could be used to improve the design of wearable robots for the hand.

Comments

Do you want to comment on this paper? The comments will show up here and if appropriate the comments will also separately be forwarded to the authors. You need to login/create an account to comment on articles. Click here to login/create an account.